Description
MI-20 Infusion Pump by SIRUSMED.
Technical Specifications
| Volume Infused | 0.01–9999.99 ml, Step by 0.01 ml |
| Flow Rate | 0.01–1800 ml/h, Step by 0.01 ml/h |
| Accuracy | ±5% |
| Purge/Bolus Rate | 1–1800 ml/h adjustable, ±20% |
| KVO Rate |
Flow rate ≥10 ml/h: KVO rate 3 ml/h Flow rate <10 ml/h and ≥1 ml/h: KVO rate 1 ml/h Flow rate <1 ml/h: KVO rate = Flow rate |
| Alarm |
Low battery, Battery exhausted, Infusion near end alarm, Air in line, Occlusion, Door open, Pause overtime, VTBI completion, Malfunction, A.C. and D.C. Off |
| Occlusion Pressure |
Three levels: High, Medium, Low H: 120 kPa ±30 kPa M: 80 kPa ±20 kPa L: 40 kPa ±20 kPa |
| Power Adapter |
A.C. input power: 100V–240V / 50–60Hz D.C. output power: 15V / 2A |
| Lithium Battery | Nominal Voltage: 10.8V; Device can work over 6 hrs on battery |
| Infusion Pump Input Power | D.C. 15V |
| Power | <55VA |
| Classification | Class II, Type CF, IPX2 |
| Weight | 1.5 kg |
| Dimension | 215 × 129 × 80 mm (W × D × H) |





Reviews